Envirco NSF Biohazard Cabinet Manuel d'utilisateur Page 1

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Équipement Envirco NSF Biohazard Cabinet. Envirco NSF Biohazard Cabinet User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1 - BIOHAZARD

Biohazard Cabinet Manual NSF Std 49 (92) 011130BIOHAZARDCABINETMANUALModel 10448 - EBC 47, 4-Foot Type A/B3Model 10276, 6-Foot Type AModel 10451, 6-Fo

Page 2 - WARNING!

10Airflow10448 - EBC 47, 4-Ft Type A/B3:Direct Inflow Method Type A/B3-Velocity 100 – 110 fpm-Airflow 253 – 278 cfmCalculated Face Velocity 115 – 125

Page 3

11NOTES:1. In performing the velocity profile test, the front row of velocity readings(behind the window) must be taken with the thermoanemometer pro

Page 4 - TABLE OF CONTENTS

12VII. INSTALLATIONCabinet LocationThe cabinet should be placed in an area of the laboratory free from strong aircurrents, heavy traffic and particul

Page 5 - II. DEFINITIONS

13Service ConnectionMake sure that a 20 amp, 115 V outlet with the proper receptacle is available for the10448 - EBC 47, 4-Ft Type A/B3. Both the 102

Page 6 - III. DESCRIPTION

14External Exhaust Connection (Optional)Under normal conditions, the Class II Type A Cabinet does not require externalexhaust from the laboratory. Ho

Page 7 - IV. FEATURES

1510” diameter centered on thimbleTop View10451, 6-Ft Type B3Top View10448, 4-Ft Type A/B3Figure 6: Thimble Exhaust as Viewed from Front of Unit20.87”

Page 8 - CABINET DIMENSIONS

16VIII. TESTING AND CERTIFICATIONThe ENVIRCO Biohazard Cabinet has been fully tested by the manufacturer prior toshipment, however, shipping and hand

Page 9 - VI. SPECIFICATIONS

17Re-certificationIt is recommended that the ENVIRCO Biohazard Cabinet be re-certified every six(6) months and the HEPA filters are replaced if necess

Page 10

18Testing1. Calibrate the photometer with the aerosol generator so that the upstreamconcentration is 100 percent. The upstream concentration for all

Page 11

19Drawing NumberPD – 10448-EBC 47WIRING DIAGRAM SIDE SECTION FRONT ELEVATION (Access Panels Removed for Clarity)PICTORIALIX

Page 12 - VII. INSTALLATION

2CAUTIONS AND WARNINGSREAD AND SAVE THESE INSTRUCTIONSWARNING!TO REDUCE THE RISK OF FIRE, ELECTRICAL SHOCK, OR INJURY TO PERSONS,OBSERVE THE FOLLOWING

Page 13

20Drawing NumberPD – 10276 / 10451WIRING DIAGRAM SIDE SECTION FRONT ELEVATION (Access Panels Removed for Clarity)PICTORIA

Page 14

21XI. FILTER REPAIRSCaution: If the cabinet has been used with infectious agents, decontaminationis necessary before removing access panels.Shipping

Page 15

22XII. AIRFLOW ADJUSTMENTAirflow adjustments can be made with either the solid state speed control inside theelectrical control panel and/or with the

Page 16

23Rear of work deckProbe set at the bottom edge of the windowFigure 9: Downflow Velocity Grid, Models 10276, 6-Ft Type A and 10451,6-Ft Type B3Inflo

Page 17 - NSF 49 (1992 edition)

24Exhaust Velocity Grid10448 EBC 47, 4-Ft Type A/B3 Exhaust Velocity Grid10276, 6-Ft Type A10451, 6-Ft Type B3Figure 10: Exhaust

Page 18

25Electrical TestsElectrical grounding tests should be made to assure that all exposed metal surfacesof the cabinet are grounded.Equipment0-1 ohm full

Page 19 - PICTORIAL

26Surface MaintenancePainted surfaces and the outside of the safety viewscreen may be cleaned withsoap and water. A light abrasive soap can be used o

Page 20 - PD – 10276 / 10451

27Working in the Biohazard CabinetCaution: Because of the highly specialized nature of the ENVIRCO BiohazardCabinet, it is recommended that an Authori

Page 21 - XI. FILTER REPAIRS

28XV. CABINET PRESSURE TESTINGThe ENVIRCO Biohazard Cabinet is designed so that there are no contaminatedplenums under positive pressure when the cab

Page 22 - XII. AIRFLOW ADJUSTMENT

29XVI. LABELING USED ON THE CABINETReview all labeling on the ENVIRCO Biohazard Cabinet as it provides both theinstaller and the worker with importan

Page 23 - (Typical)

3E. Caution: Do not, under any circumstances, cut or remove the grounding prong from theplug.5-20 PPlug5-20 RReceptacle5-30 PPlug5-30 RReceptacle1044

Page 24 - Exhaust Velocity Grid

30XVII. WIRING DIAGRAMSFigure 11: 10448 - EBC 47 (4-Ft Type A/B3) Wiring DiagramM1 6 5 4 3 2 1 1 2 3 3 AMPS MAXCONN 'G'

Page 25

31Figure 12: 10276 (6-Ft Type A) and 10451 (6-Ft Type B3) Wiring DiagramM1 6 5 4 3 2 1 1 2 3 3 AMPS MAX'E1' 'C&apos

Page 26

32XVIII. HEPA FILTER REPLACEMENTThe HEPA filters require replacement when the air resistance is such that thespecified air velocities cannot be achie

Page 27

33XIX. REPLACEMENT AND OPTIONAL PARTSPART NUMBERDESCRIPTION 10448 – EBC 47 10276 / 10451Assembly, Motor/Blower 23002 22425 (2 ea.)Motor, 3/4 H.P. O.D

Page 28 - XV. CABINET PRESSURE TESTING

34XX. BIBLIOGRAPHY1. Effective Use of the Laminar Flow Biological Safety Cabinet. U.S. Departmentof Health, Education and Welfare, Office of Biohaz

Page 29 - FLAMMABLE

35XXI. GLOSSARYAerosol: A colloid of liquid or solid particles suspended in a gas, usually air.Ambient: Surrounding on all sides.Biohazard: A cont

Page 30 - XVII. WIRING DIAGRAMS

36XXII. RISK LEVELS, CLASSIFICATIONLow RiskThe risk level of the agents and/or the operation has minimal effect on personnel,animals, or plants under

Page 31

37XXIII. MANUFACTURER'S LIMITED WARRANTYThe Biohazard Cabinet is manufactured by ENVIRCO, of Albuquerque, New Mexico andis warranted to be free f

Page 32

4TABLE OF CONTENTSCAUTIONS AND WARNINGS...2TABLE OF CO

Page 33 - Fax : (505) 345-8875

5I. INTRODUCTIONInvestigators and technicians concerned with microbiological safety have for manyyears utilized specialized containment enclosures to

Page 34 - XX. BIBLIOGRAPHY

6Class II, Type A/B3 Cabinets:1. Meets the requirements of Class II, Type A cabinets when the HEPA filteredexhaust air is recirculated back to the la

Page 35 - XXI. GLOSSARY

7IV. FEATURES1. The hinged viewscreen is constructed with safety glass and has a seven-degree angle for comfortable viewing. The hinge enables easy

Page 36

8V. CABINET DIMENSIONSModel A B C D E F10448 – EBC 47, 4-Ft Type A/B3 47.0” 32.5” 62.0” 58.0” 4.0” 28.5”10276, 6-Ft Type A10451, 6-Ft Type B370.0” 32

Page 37 - (800) 345-6094

9VI. SPECIFICATIONSElectrical:10448 - EBC 47, 4-Ft Type A/B3: 115v, 60 Hz, 20 amp10276, 6-Ft Type A: 115v, 60 Hz, 30 amp10451, 6-Ft Type B3: 115v, 60

Commentaires sur ces manuels